GlobeCore / Instruments - Transformer Oil Testers / Breakdown Voltage Tester TOR-100 (TOR-100 SL)

Breakdown Voltage Tester TOR-100 (TOR-100 SL)

The TOR-100 tester is designed for determining the electrical breakdown voltage of transformer oil and other liquid dielectrics, following standards such as IEC 60156, VDE 0370, ASTM D877, ASTM D1816, and IRAM 2341. It’s worth noting that this list of standards can be adapted and expanded to accommodate customer preferences and updates in standards.

Principle of Operation

The core functionality of the tester revolves around a high-voltage transformer and electronic modules. The device operates by progressively increasing the voltage applied to the secondary winding of the high-voltage transformer, starting from zero and rising to the maximum value. Alternatively, it stops at a voltage level where a breakdown occurs within a liquid dielectric contained in a measuring cell. This cell is installed on the high-voltage contacts of the transformer and serves as an integral part of its output electrodes.

The elevation of voltage at the high-voltage contacts in the secondary winding of the transformer is achieved through a gradual increase in voltage on the primary winding. This primary winding receives a sinusoidal voltage from a controlled master inverter.

No.

Parameter

Value

1 Operating AC voltage, V

85 – 264

2 Power frequency, Hz

48 – 63

3 Power requirement, VA

70 max

4 Maximum output voltage, kV

Sinusoidal, up to 100 kV RMS

5 Measurement tolerance of output voltage

±1 %

6 Voltage increase rate, kV/s

From  0.5 to 10

7 Resolution of output voltage indication, V

100

8 High voltage shutoff time after breakdown, µs

10 max,

4 typical

9 Measuring cell volume, cm³

500

10 Ambient temperature measurement range, °С (°F)

-10 – 50 (14 – 212)

11 Resolution of temperature measurement, °С

1

12 Integrated printer

Yes

13 Operating temperature, °С (°F)

-10 – 50 (14 – 122)

14 Storage temperature, °С

From – 20 to + 60

15 Relative humidity, %

Up to 90 without condensation

16 Atmospheric pressure, mm Hg

630 – 800

17 Dimensions, mm

510х300х350

18 Weight, kg, not more than

30

  • Automated and precise measurement of breakdown voltage, capable of reaching up to 100 kV.
  • Immediate (4 µs) test voltage shutdown upon breakdown occurrence.
  • Equipped with an integrated printer for convenient printing of measurement results.
  • Includes a USB connector for connecting a flash drive, allowing the archiving of test data for later transfer to a computer.
  • Simple and user-friendly in terms of operation and maintenance.

    GlobeCore

    Leave your request